• MK SPORTS集团(中国)

    Resource 4: rEV Serial Dilution

    Joshua A Welsh, Sean M Cook, Jennifer Jones, Joanne Lannigan, Vera A. Tang

    Published: 2023-11-02 DOI: 10.17504/protocols.io.kxygx9opzg8j/v1
    AI 解读
    浏览 1356

    Disclaimer

    This protocol summarizes key steps for a specific type of method, which is one of a collection of methods and assays used for EV analysis in the NCI Translational Nanobiology Section at the time of submission of this protocol. Appropriate use of this protocol requires careful, cohesive integration with other methods for EV production, isolation, and characterization.

    Abstract

    Flow cytometry (FCM) is a common extracellular particles (EPs), including viruses and extracellular vesicles (EVs), characterization method. Frameworks such as MIFlowCyt-EV exist to provide reporting guidelines for metadata, controls, and

    data reporting. However, tools to optimize FCM for EP analysis in a systematic and quantitative way are lacking. Here, we demonstrate a cohesive set of methods and software tools that optimize FCM settings and facilitate cross-platform comparisons for EP studies. We introduce an automated small particle optimization (SPOT) pipeline to optimize FCM fluorescence and light scatter detector settings for EP analysis and leverage quantitative FCM (qFCM) as a tool to further enable FCM optimization of fluorophore panel selection, laser power, pulse statistics, and window extensions. Finally, we demonstrate the value of qFCM to facilitate standardized cross-platform comparisons, irrespective of instrument configuration, settings, and sensitivity in a cross-platform standardization study utilizing a commercially available EV reference material.

    Steps

    rEV Reconstitution

    1.

    Briefly centrifuge 100x g,4°C rEVs before opening.

    2.

    Add 100µL of 4°C deionized water.

    3.

    Pipet up and down to mix.

    Note
    Do not vortex

    4.

    Aliquot 20µL into low-binding Eppendorf tubes.

    Equipment

    ValueLabel
    Low Protein Binding Microcentrifuge TubesNAME
    Microcentrifuge TubesTYPE
    Thermo ScientificBRAND
    90410SKU
    5.

    Store at 2-4°C for up to one day (up to 2 weeks at -80°C).

    rEV Dilutions

    6.

    Label 10 low-binding Eppendorf tubes:

    AB
    LabelDilution
    150
    2100
    3200
    4400
    5800
    61600
    73200
    86400
    912800
    1025600
    7.

    Add 980µL of DPBS (preferably 0.1µm filtered) to the tube 1 from .

    8.

    Add 500µL of DPBS (preferably 0.1 µm filtered) to the remaining tubes.

    9.

    Add 20µL of the stock reconstituted rEVs to the tube labeled 1 ( ), pipet up and down to mix.

    Note
    Do not vortex

    10.

    Pipette 500µL of the 1:50 dilution in tube 1 into the 1:100 dilution in tube 2, pipet up and down to mix.

    Note
    Do not vortex

    11.

    Repeat the above step for each of the remaining dilution tubes until the serial dilution has been completed for all dilutions.

    12.

    Keep on ice during acquisition.

    Flow Cytometer Acquisition

    13.

    Use the instrument settings that were established from the previous gain incrementation and scatter calibration resources. On the Aurora, set the window extension to 0. On the CytoFLEX, use the high acquisition mode. Collect for at least 1 minute at a low flow rate with a 30 second recording delay if using plate mode.

    Note
    To set the window extension to 0 on the Cytek Aurora, navigate to the 'Lasers' tab under the sample acquisition settings and set the window extension to 0.To change the acquisition mode on the CytoFLEX S, click on the 'Advanced' menu on the top-of-the-screen ribbon. Next click on 'Event Rate Settings' menu, change the acquisition mode to 'High', and press 'Ok'.

    13.1.

    Set your instrument to trigger on the most sensitive side scatter channel at the FCMPASS optimal gain output as determined by the FluoSpheres scatter voltration. Set the threshold in the noise such that the event rate is ~1000 events/sec when running DPBS.

    Resource 1: Scatter Detector Setting Incrementation for FCMPASS

    13.2.

    Set the FITC parameter to the FCMPASS optimal gain output as determined by the QbSure fluorescent voltration.

    Resource 2: Fluorescence Detector Setting Incrementation for FCMPASS

    14.

    Collect a DPBS sample in a FACS tube for one minute on low.

    Note
    This is to ensure the instrument is clean on the side scatter and fluorescent channels before acquiring the rEVs.

    DPBS acquired on the Cytek Aurora and CytoFLEX S.
    DPBS acquired on the Cytek Aurora and CytoFLEX S.
    15.

    Collect all rEVs samples for one minute on low, starting with the highest dilution (least concentrated) and working down to lowest dilution (most concentrated). This will reduce the effect of carry over from high concentrations.

    rEVs acquired on the Cytek Aurora and CytoFLEX S.
    rEVs acquired on the Cytek Aurora and CytoFLEX S.
    16.

    Run a DPBS sample in a FACS tube to make sure there are no carryover events; there should be no events in the B1 and B2 channel.

    Note
    If there is carry over clean instrument with bleach and flush with deionized water as appropriate.

    17.

    Label as FACS tube as 'QbSure', and add 500µL DPBS. Vortex the QbSure beads for 5 sec and add 3 drops QbSure beads into FACS tube. Run QbSure beads on low and collect 10,000 bead events.

    QbSure beads gated on blue SSC-H and FSC-H on the Cytek Aurora and CytoFLEX S. The violet SSC settings remain unchanged. In order to limit .fcs file size, a FSC scatter trigger may be used until 10000 QbSure bead events are acquired.
    QbSure beads gated on blue SSC-H and FSC-H on the Cytek Aurora and CytoFLEX S. The violet SSC settings remain unchanged. In order to limit .fcs file size, a FSC scatter trigger may be used until 10000 QbSure bead events are acquired.

    Note
    You may need to adjust the B-SSC and FSC detector settings for the bead acquisition. Do not adjust fluorescence detector settings or high sensitivity SSC detector settings. Assuming the QbSure beads were cross calibrated in the previous protocols, this allows for the calibration of the rEV fluorescent data into calibrated units.

    18.

    Run the 100 nm polystyrene NIST beads at the same settings as the rEVs (diluted in DPBS to a concentration of 5E6 p/mL) until at least 10000 bead events have been acquired.

    Note
    Acquiring the 100 nm polystyrene NIST bead allows for the calibration of the rEV side scatter data into calibrated units if the collection half-angle of the instrument has been determined in previous protocols.

    19.

    If calibration of data into standard units is desired, the 100 nm polystyrene NIST bead and QbSure beads can be used to calibrate data by following the FCMPASS experiment calibration protocols.

    FCMPASS Protocol Collection

    推荐阅读

    Nature Protocols
    Protocols IO
    Current Protocols
    扫码咨询